Unmasking of atrial repolarization waves using a simple modified limb lead system
1Department of Instrumentation and Control Engineering, National Institute of Technology; Tiruchirappalli-India
2Department of Cardiology, Madras Medical College, Rajiv Gandhi Government General Hospital; Chennai-India
Anatol J Cardiol 2015; 8(15): 605-610 PubMed ID: 25550180 PMCID: 5336858 DOI: 10.5152/akd.2014.5695
Full Text PDF

Abstract

Objective: In the present study, a modified limb lead (MLL) system was used to record the Ta wave in sinus rhythm and with AV block in male patients. Materials and Methods: Eighty male subjects (mean age 36±7 years) in sinus rhythm and 20 male patients with AV block (mean age 72±5 years) were included in this study. Standard limb lead (SLL) ECGs and MLL ECGs were recorded for 60 seconds each with an EDAN SE-1010 PC ECG system. Results: In sinus rhythm subjects, the observable Ta wave duration was 109±4.7 ms, the P-Ta duration was 196±5.1 ms, and the corrected P-Ta duration was 238±7.2 ms. The Ta wave peak amplitude was -42±8 µV. In AV block patients, the Ta wave duration was 314±28 ms the P-Ta duration was 418±29 ms and the corrected P-Ta duration was 462±31 ms, while the Ta wave peak amplitude was -37±9 µV. A correlation was found between the P and Ta wave amplitude, and no correlation was found between the P and Ta wave duration or the Ta amplitude and Ta duration in sinus rhythm and AV block subjects. Conclusion: The end of the Ta wave is not observable in sinus rhythm subjects, as it extends into the QRS complex and ST segment. In AV block patients, the Ta wave duration was generally three times longer than the observable Ta duration in sinus rhythm subjects.


Basit modifiye bir ekstremite elektrot sistemi kullanılarak atriyal repolarizasyon dalgalarının ortaya konması
1Department of Instrumentation and Control Engineering, National Institute of Technology; Tiruchirappalli-India
2Department of Cardiology, Madras Medical College, Rajiv Gandhi Government General Hospital; Chennai-India
The Anatolian Journal of Cardiology 2015; 8(15): 605-610 DOI: 10.5152/akd.2014.5695 PMID: 25550180

Amaç: Bu çalışmada, bir Modifiye Ekstremite Elektrot (MEE) sistemi erkek hastalarda sinüs ritminde ve AV bloklularda Ta dalgasını kaydetmek için kullanıldı. Yöntemler: Sinüs ritminde 80 erkek gönüllü (ortalama yaş 36±7 yıl) ve AV bloklu 20 erkek hasta (ortalama yaş 72±5 yıl) bu çalışmaya dahil edildi. Standart Ekstremite Elektrotlu (SEE) EKG’ler ve MEE EKG’ler bir EDAN SE-1010 PC EKG sistemi ile her biri 60 saniye süreyle kaydedildi. Bulgular: Sinüs ritmindeki gönüllülerde gözlemlenebilir Ta dalgası süresi 109±4,7 ms, P-Ta süresi 196±5,1 ms ve düzeltilmiş P-Ta süresi 238±7,2 ms idi. Ta dalgası tepe amplitüdü -42±8 µV idi. AV bloklu hastalarda, Ta dalgası süresi 314±28 ms, P-Ta süresi 418±29 ms ve düzeltilmiş P-Ta süresi 462±31 ms iken Ta dalga tepe amplitüdü -37±9 µV idi. Sinüs ritimli ve AV bloklu kişilerde P ve Ta dalga amplitüdü arasında korelasyon bulundu, P ve Ta dalga süresi veya Ta amplitüdü ve Ta süresi arasında korelasyon saptanmadı. Sonuç: QRS kompleksi ve ST segmenti içine uzandığı için Ta dalgasının sonu sinüs ritimli kişilerde gözlemlenebilir değildir. AV bloklu hastalarda Ta dalga süresi, sinüs ritimli kişilerde gözlemlenebilir Ta süresinden genellikle üç kat daha uzundu.