Hemodynamic performance of mechanical aortic valves in narrow aortic annulus cases [Anatol J Cardiol]
Anatol J Cardiol. 2005; 5(1): 30-33

Hemodynamic performance of mechanical aortic valves in narrow aortic annulus cases

Aytül Belgi1, Seyhan Çetin1, Emre Altekin1, Bekir Kalaycı1, Selim Yalçınkaya2, Mehmet Kabukçu1, Oktay Sancaktar1, Filiz Ersel Tüzüner1
1Akdeniz University, Medical Faculty, Department of Cardiology, Antalya
2Akdeniz Üniversitesi Tıp Fakültesi Kardiyoloji Anabilim Dalı, Antalya, Türkiye

Objective: Small-sized prosthetic valves used in aortic valve position usually cause a high transvalvular gradient. This situation, especially in people with a high body surface area, causes a patient-prosthetic valve mismatch. In this study, hemodynamic performances of the two new generation, bileaflet, small size mechanical valves were evaluated during rest and stress conditions. Methods: Eighteen patients with implanted 21-mm St. Jude HP (11 cases) and Sorin Bicarbon (7 cases) prosthetic valves in aortic position were evaluated at rest and during dobutamine infusion, 16.3±5.5 months after the operation. Dobutamine infusion was started at a dose of 5 mg/kg/min. The infusion rate was increased every 15 minutes up to 10-20-30 mg/kg/min. Maximum gradient, mean gradient, effective valve area (EVA), effective valve area index (EVAI) and performance index (PI) were calculated during rest and maximum dobutamine infusion rate. Results: When the rest and dobutamine infusion measurements were compared, the maximum and mean gradients were found to be increased moderately, but EVA, EVAI, PI values did not change with increasing heart rates. For St. Jude valve; maximum gradient increased from 30.5±6.5 mmHg to 74.7±33.6 mmHg (p=0.03) and mean gradient increased from 17.9±3.8 to 40.8±23.8 mmHg (p=0.03). For Sorin Bicarbon valve; maximum gradient increased from 31.7±13.3 mmHg to 52.0±11.2 mmHg (p=0.01), mean gradient increased from 16.1±6.4 mmHg to 28.8±1.0 mmHg (p=0.01). The difference was not significant between the two valves with respect to measured parameters during rest and maximum dobutamine infusion. Conclusion: According to our findings, 21-mm Sorin Bicarbon and St. Jude HP mechanical bileaflet valves have good hemodynamic performance during exercise and these types of valves seem to be appropriate for patients with small aortic annulus.

Keywords: Aortic valve replacement, dobutamine stress echocardiography

Aytül Belgi, Seyhan Çetin, Emre Altekin, Bekir Kalaycı, Selim Yalçınkaya, Mehmet Kabukçu, Oktay Sancaktar, Filiz Ersel Tüzüner. Hemodynamic performance of mechanical aortic valves in narrow aortic annulus cases. Anatol J Cardiol. 2005; 5(1): 30-33
Manuscript Language: English


Journal Metrics

Journal Citation Indicator: 0.18
CiteScore: 1.1
Source Normalized Impact
per Paper:
0.22
SCImago Journal Rank: 0.348

Quick Search



Copyright © 2024 The Anatolian Journal of Cardiology



Kare Publishing is a subsidiary of Kare Media.